ISR Quick Reference Guide


Patients must be loaded from the side of table.
Tilting or instability may occur if patient is loaded onto the pedestal end of table or the imaging end of the table.
SETUP/PATIENT PREPARATION
Moving table
Never move the table with a patient onboard
- Unlock Rear Casters using Caster Lock Pedal.
- Unlock Front Steering using Front Steer Lock Pedal.
- Position table to desired location.
- Secure table mobility by locking Casters and front steering using Rear Caster Lock Pedal and Front Steer Lock Pedal.
AC Power
- Plug power cord into properly grounded outlet.
- Protect power cord from damage by casters or other equipment.
Controls
- Emergency Stop Pushbutton disables all tabletop motion control when engaged.
- If tabletop motion controls fail to operate, make sure that the Emergency Stop Pushbutton is NOT pushed in.
Battery Operation (DC Power)
- The table is FULLY functional, but motions are slower.
- Connect to AC power for a minimum of 8 hours daily to ensure a proper battery charge.
- Failure to push in Emergency Stop Pushbutton when table is not in use and not connected to AC power for longer than 72 hours, may result in permanent damage to the batteries.
CPR
Return table to level, retract tabletop to minimize overhang and lower table position to a comfortable height to perform CPR.
Brakes
The ISR table braking system utilizes a passive magnetic brake; it remains engaged in the event of a complete loss of electrical power. Power is necessary to release brakes.
Braking force is limited by the magnet strength. Use caution when applying pressure to the side of the table as the brakes may be overcome if more than 50 pounds of force is applied.
To reduce potential motion during patient transfers with larger patients, release the table brake and move the top in the direction the patient will be pulled during loading or unloading until the tabletop transverse travel limit is reached. Release the panning handle to re-engage the lock and then transfer the patient.
Cleaning and storage
- Unplug power cord and wrap it around the cord holder.
- Push in Emergency Stop Pushbutton.
- Lock rear casters and front steering.
- Clean and disinfect table.
Table Locking
Table Motion Control Panhandle
Caster Lock & Steer Pedal Postions
Emergency Stop Pushbutton